Class II Biological Safety CabinetsClass II biological safety cabinets provide personnel,
environmental and product protection. Air drawn into the cabinet passes
through a high efficiency particulate air (HEPA) filter before it passes
over the work surface. Class II biological safety cabinets
are further broken down into types A1, A2, B1, and B2.
The class IIA cabinet recirculates approximately 70% of its air flow within the cabinet. This cabinet exhausts into the room or through the building exhaust by a canopy unit. (See Figure 1 and Figure2)
(Can be room re-circulated or air gap type (thimble).)
(Can be room re-circulated or air gap type (thimble).) Class II Type B1 cabinet recirculates approximately 30% of its
air flow within the cabinet. This cabinet exhausts to the outside through
a duct system. (See Figure 3)
(Hard-ducted.)
The class II Type B2 cabinet does not recirculate any air within
the cabinet. All exhaust leaves the cabinet through a duct to the
outdoors. (See Figure 4)
(Hard-Ducted.)